221. The sensitivity of an ultrasonic test system |
a. depends on the search unit, pulser, and amplifier used |
b. decreases as the frequency is increased |
c. increases as the resolution increases |
d. Is not related to mechanical damping or the transducer |
222. The ability of a test system to separate the front-surface echo and the echo from a small discontinuity |
just below the surface |
a. depends primarily upon the pulselength generated from the instrument |
b. is not related to the surface roughness of the part under inspection |
c. is primarily related to the thickness of the part under inspection |
d. is usually improved by using a larger diameter search unit |
223. Search unit sensitivity is most often determined by |
a. calculations based on frequency and thickness of piezoelectric element |
b. the amplitude of the response from an artificial discontinuity |
c. comparing it to a similar search unit made by the same manufacturer |
d. determining the rinsing time of search unit |
224. Notches are frequently used as reference reflectors for |
a. distance amplitude calibration for shear waves |
b. area amplitude calibration |
c. thickness calibration of plate |
d. determining near-surface solutions |
225. Notches provide good reference discontinuities when UT examination Is conducted to primarily |
defects such as |
a. porosity in rolled plate |
b. inadequate penetration at the root of a weld |
c. weld porosity |
d. internal inclusions |
226. The difference between a compression and shear wave is |
a. quantitative measure |
b. particle direction |
c. qualitative measure |
d. amplitude |
227. The particle motion for Rayleigh waves is usually described as |
a. sinusoidal |
b. circular |
c. elliptical |
d. shear |
228. Based upon wave theory and ignoring attenuation losses, the echo amplitude is |
a. directly proportional to the distance to the reflector |
b. inversely proportional to the distance to the reflector |
c. directly proportional to the square of the distance to the reflector |
d. inversely proportional to the square of the distance from the reflector |
229. The rate generator In B-scan equipment will invariably be directly connected to the |
a. CRT Intensity circuit |
b. pulser circuit |
c. RF amplifier circuit |
d. horizontal sweep circuit |
230. In A-scan equipment, the RF pulser output voltage is normally in the range of |
a. 1-10 V |
b. 10-100 V |
c. 100-1000 V |
d. 1000-3000 V |
231. When contact testing. an increase In tightness of a shrink fit to a hollow shaft will cause the ratio |
of the back reflection to the metal-to-metal interface reflection to |
a. increase |
b. decrease |
c. remain unchanged |
d. cannot be predicted as the response is material-dependent |
232. During the Inspection of a parallel-sided, machined forging using the straight beam immersion technique |
a diminished back reflection in a localized area in the absence of a defect indication would least |
likely represent |
a. a coarse grain structure |
b. small nonmetallic stringers |
c. a defect oriented at a severe angle to the entry surface |
d. a large Inclusion |
233. As you increase the frequency of a given size transducer |
a. the lateral resolution is impaired |
b. the beam divergence increases |
c. the length of the near field increases |
d. the sensitivity increases |
234. Which frequency ran best distinguish the difference between a large planar defect and 4 stacked |
(multiple-layered) laminations in rolled plate? |
a. 0.5 MHz |
b. 1 MHz |
c. 2.25 MHz |
d. 5 MHz |
235. During immersion examination. when evaluating the response from a contoured surface of a part |
irrelevant indications due to the contour are most likely to appear as |
a. sharp spiked signal Indications |
b. irregular signal indications |
c. broad-based signal Indications |
d. hash |
236. The pulse applied to the electrodes of the ultrasonic search unit is |
a. electrical |
b. mechanical |
c. electro-mechanical |
d. piezoelectrical |
237. In calibrating at ultrasonic test instrument utilizing the responses from each of the area amplitude |
type reference blocks, the determination of the |
a. vertical range is obtained |
b. Pulse range is obtained |
c. resolving range is obtained |
d. horizontal range is obtained |
238. Test sensitivity corrections for metal distance and discontinuity area responses are accomplished by using |
a. an area amplitude set of blocks |
b. an area amplitude and a distance amplitude set of blocks |
c. a distance amplitude set of blocks |
d. steel balls of varying diameter |
239. The time from the start of the ultrasonic pulse until the reverberations completely decay limits the |
maximum usable |
a. pulse time-flaw rate |
b. pulser/receiver rate |
c. pulse repetition rate |
d. modified pulse-time rate |
240. Rough surfaces can cause undesirable effects which are noticeable when parts are tested ultrasonically |
including |
a. annular maxima rings |
b. an increase in the width of front face echo and consequent loss of resolving power |
c. acoustical mismatch |
d. asymmetrical modes |
answer
221 | 222 | 223 | 224 | 225 | 226 | 227 | 228 | 229 | 230 |
A | A | B | A | B | B | C | D | B | A |
231 | 232 | 233 | 234 | 235 | 236 | 237 | 238 | 239 | 240 |
A | D | C | D | C | A | A | B | C | B |